A very common use of the forces on a coil in a magnetic field is that of
the moving coil meter, shown in Figure 1.
The coil is suspended between the poles of a
magnet on jewelled bearings and is held in place by two finely coiled springs (S1 and S2) through
which the current to be measured passes in and out of the coil.
The pole pieces are
shaped so that the magnetic field is radial thus giving the maximum and constant torque on the
coil whatever its position (Figure 2). There is a soft iron armature in the centre of the coil, and
this further concentrates the magnetic field through it due to the high value of the relative
permeability of the material of the core.
The coil is supported on a light metal frame and
induced currents in this frame give electromagnetic damping of the movement.
When it is in
equilibrium with a current passing through it, the torque on the coil produced by the magnetic
field is balanced by an opposing torque due to the rigidity of the springs. Clearly the more
delicate the springs the bigger the deflection for a given current.
Some meters are of the
suspended coil type; the coil is mounted on a fine phosphor bronze wire and the deflection
measured with a lamp and scale. A small mirror mounted above the coil reflects a beam of light
on to a scale, the angle of twist of the light beam being double the angle of rotation of the coil.
Such instruments have a framework of brass or aluminium.
Meters are often made with
a protective series resistor since too large a current will burn out the springs or the coil. When
the correct range for the meter has been found this protective resistor may be shorted
out.
The moving coil analogue meter is used less and less these days having been
replaced in many applications by the digital
meter.
The amount of
deflection of the pointer in a meter for a certain current depends on the design of the meter. The
amount of angular twist for a unit (1A) current is called the current sensitivity of the
meter.
Consider a coil of N turns and cross-sectional area A, carrying a current I in a
field of flux density B. The torque C due to the magnetic field is given
by
To
increase this (that is, to make the meter more sensitive) we require:
(a) large magnetic flux
density (B), that is, the gap between the poles as small as possible,
(b)a coil with a large
area (A),
(c) a large number of turns (N), and
(d) a small value of k that is, a very thin
wire or one with a very low rigidity.
Unfortunately (b) and (c) tend to make the coil both
bigger and heavier and so cause problems with (a) and (d). A compromise has to be
reached.
The radial field is important since for this arrangement B sinq is constant and
therefore the angle of twist q is directly proportional to the current I for all positions of the coil.
A typical moving coil meter such as those used in schools may have a full
scale deflection (fsd) of 100 mA and a resistance of 1000
W. This means that the pointer will deflect right across the
scale when a current of 100 mA is passed through the
meter.
When this occurs the p.d. between the terminals of the meter will be 100 mA x 1000 W = 100 mV.
Clearly this is quite inadequate when measurements of currents of say 5 A or voltages of 12 V
are required. External resistors may be used to extend the range of the meter and these are
known as shunts and series resistors.
So that as little current as possible is drawn from a
circuit under test a good voltmeter should have a resistance of at least 1000W per volt - for example a meter designed to read. voltages up to
10 V should have a resistance of 10 kW. This also requires
a galvanometer with a high current sensitivity
The moving coil galvanometer can also be
used as an ohmmeter and a wattmeter.
There has been a considerable increase in the
use of direct-reading digital meters in the last few years. These rely on a totally different
principle, that of the integrated circuit, for their operation and have no moving parts. The digital
voltmeter has a very high resistance (of the order of 10 MW
on d.c.) but it does need a small internal battery to power the instrument. The input voltage to be
measured is compared with a steadily rising voltage produced by a ramp generator (an
application of the op amp). The time taken for the rising voltage to reach that of the input voltage
is measured and this time is directly proportional to the voltage. The output reading is scaled to
give a direct reading in volts.