When X-rays pass through matter such as a
human body they will lose energy in one or more of the following ways:
(a)
Scattering – the energy of the X ray photon is not sufficient to cause electron emission from
the atom (see Compton scattering) from 1 to 30 keV
(b) The photoelectric effect - an X-
ray photon transfers all its energy to an electron which then escapes from the atom from 1 to
100 keV
(c) Compton scattering - an X-ray photon collides with a loosely-bound
outer electron. At the collision the electron gains some energy and a scattered X-ray photon
is produced travelling in a different direction from the incident photon and with a lower energy
from 0.5 to 5 MeV
(d) Pair production - an X-ray photon with an energy greater than
1.02 MeV enters the intense electric field at the nucleus. It may be converted into particles, a
positron and an electron. These two particles usually annihilate each other producing two
photons, each of energy 511 keV – above 5 MeV.
In a vacuum the intensity (I) of
the X ray beam will fall off as the inverse of the square of the distance (d) from the
source.
The graph shown in Figure 1 is for an X beam of one energy - monoenergetic. In this case the half value thickness of the
material is always the same no matter how much the original intensity has been reduced. If
the beam has a spread of energies then the further through the material the X ray beam
passes the longer the half value thickness becomes.
This is because the lower
energy X rays in the beam are absorbed more easily than the higher energy ones.
This
is shown in Figure 2. The X ray beam contains a greater proportion of higher energy X rays
the further into the material it penetrates. This is said to be a 'hardening' of the
beam.
Therefore as the beam goes deeper into the material the 'average' energy of the X rays in the beam increases and so a greater thickness of absorber is needed to further reduce the beam intensity by a half.
Figure 3 shows the graph for a beam of X rays with a spread of energy. It shows that the absorption curve
flattens as the beam penetrates further into the material and the half value thickness also
increases.
The equivalent energy of an X ray beam is shown by its half value thickness in a particular material. For
example X rays from an X ray tube with a peak anode voltage of 80 kV have a half value
thickness of about 3 mm of aluminium. However the half value thickness for a 28 keV
monoenergetic beam of X rays is 3 mm of aluminium and so the equivalent energy of the 80
kV X ray beam is 28 keV.
The quality of the beam is determined by the spread of X ray
energies across the spectrum